Engineered Water Systems Changsha · China info@kertonewater.com Tel: +86 193 7519 4365
Engineered Water Equipment

KZ-XDA Laboratory Ultrapure Water System Series (60/100/150 L/h)

KZ-XDA laboratory ultrapure-water family with 60, 100 and 150 L/h models, RO and UP outlets, and documented UP resistivity of 18.2 MΩ·cm at 25°C.

The KZ-XDA family combines RO and UP outlets in three source-listed capacities for laboratory workflows. This page preserves the exact model designations and reports only specifications visible in the supplied product archive.

Select specification

Changing the model updates the specifications below and the model attached to your enquiry.

ModelKZ-XDA-60L
Rated output60 L/h
Power150 W
Dimensions (L×W×H)27×60×62 cm

Technical Specifications

Model KZ-XDA-60L
Rated output 60 L/h
Power 150 W
Dimensions (L×W×H) 27×60×62 cm
Power supply AC 220 V / 50 Hz
Feed-water pressure 0.15–0.6 MPa
Operating pressure 0.4–0.6 MPa
UP resistivity 18.2 MΩ·cm at 25°C
UP TOC <10 ppb
UP heavy-metal ions <0.1 ppb
UP microorganisms <1 cfu/ml
RO conductivity N/A — the supplied artwork omits the numeric value

Model Comparison

Model Rated output Power Dimensions (L×W×H)
KZ-XDA-60L 60 L/h 150 W 27×60×62 cm
KZ-XDA-100L 100 L/h 300 W 40×61×93 cm*
KZ-XDA-150L 150 L/h 300 W 52×61×133 cm

Dimensions marked with * differ between the source parameter table and the dimension graphic — see Source Notes below.

System Design & Functions

  • Separate RO and UP dispensing outlets support general purified-water and ultrapure-water workflows.
  • LCD operating-status display and online resistivity monitoring are shown in the supplied product material.
  • Automatic RO-membrane flushing and multi-stage filtration are documented system functions.
  • Electrophoretic spray-finished cabinet, terminal filter and a UV-sterilization arrangement are shown in the source material.
  • The source graphics conflict on tank construction: one calls it ABS and another calls it 304 stainless steel. Tank material must be confirmed in the final technical agreement.

Applications

  • Electrochemistry, spectrophotometry, microbiological culture and reagent preparation.
  • Cell culture, PCR, IVF, protein purification and electrophoresis.
  • Biochemistry, proteomics, genomics and immunoanalysis.

Installation & Selection

  • Confirm feed-water quality, inlet pressure, drainage, grounded AC 220 V power and service clearance before installation.
  • Select the 60, 100 or 150 L/h model from actual peak demand and duty cycle; the archive does not provide a model-selection formula.
  • Final consumables, storage arrangement, tank material and water-quality acceptance criteria must be recorded in the approved technical agreement.

Source Notes

  • The numeric RO-conductivity threshold is not stated in the supplied artwork and is therefore reported as N/A.
  • For KZ-XDA-100L, the parameter table states 40×61×93 cm while a separate dimension graphic appears inconsistent; the table value is reproduced here and must be confirmed before ordering.
  • The archive provides no named compliance standard, warranty term or product video; these items are N/A pending confirmed manufacturer documentation.