Product Systems/KY-JY Clinical Laboratory Biochemical Water System (60 / 120 / 180 / 250 / 500 L/h)
Engineered Water Equipment
KY-JY Clinical Laboratory Biochemical Water System (60 / 120 / 180 / 250 / 500 L/h)
The KY-JY clinical laboratory biochemical water system is available in 60 / 120 / 180 / 250 and 500 L/h for biochemical analyzers of different laboratory scales.
Hunan Kertone Water Group Co., Ltd.Changsha · China
Technical Specifications
Select specification
Changing the model updates the specifications below and the model attached to your enquiry.
Model
KY-JY-60KU
Maximum output
60 L/h
Dimensions
55×35×95 cm
Core configuration
Delta booster pump, one 3013-400G RO membrane, 20G pressure tank and two polishing columns
Feed water
Municipal tap water
Feed temperature
5-38°C
Power supply
AC 220V, grounded
Purification process
RO plus multi-stage polishing columns
Model
KY-JY-120KU
Maximum output
120 L/h
Dimensions
55×35×95 cm
Core configuration
Delta pump, two 3013-400G RO membranes, 20G pressure tank and two polishing columns
Feed water
Municipal tap water
Feed temperature
5-38°C
Power supply
AC 220V, grounded
Purification process
RO plus multi-stage polishing columns
Model
KY-JY-180KU
Maximum output
180 L/h
Dimensions
55×35×95 cm
Core configuration
Two 3013-600G RO membranes, 20G pressure tank and four polishing columns
Feed water
Municipal tap water
Feed temperature
5-38°C
Power supply
AC 220V, grounded
Purification process
RO plus multi-stage polishing columns
Model
KY-JY-250KU
Maximum output
250 L/h
Dimensions
64×64×155 cm
Core configuration
Stainless frame, Delta pump, one industrial-grade 4040 RO membrane and six polishing columns
Feed water
Municipal tap water
Feed temperature
5-38°C
Power supply
AC 220V, grounded
Purification process
RO plus multi-stage polishing columns
Model
KY-JY-500KU
Maximum output
500 L/h
Dimensions
64×64×155 cm
Core configuration
Stainless frame, vertical stainless feed pump, FD2-7/8 high-pressure pump, two industrial-grade 4040 RO membranes and two-stage precision filtration
Feed water
Municipal tap water
Feed temperature
5-38°C
Power supply
AC 220V, grounded
Purification process
RO plus multi-stage polishing columns
Model Comparison
Item
KY-JY-60KU
KY-JY-120KU
KY-JY-180KU
KY-JY-250KU
KY-JY-500KU
Purified water flow
60 L/h
120 L/h
180 L/h
250 L/h
500 L/h
Filtration process
Two-stage RO + polishing columns
Two-stage RO + polishing columns
Two-stage RO + polishing columns
Two-stage RO + polishing columns
Two-stage RO + polishing columns
RO membrane
One 3013-400G RO membrane
Two 3013-400G RO membranes
Two 3013-600G RO membranes
One industrial-grade 4040 RO membrane
Two industrial-grade 4040 RO membranes
Purification columns
2
2
4
6
8
Voltage / power
220V / 100W
220V / 100W
220V / 100W
220V / 100W
220V / 100W
Conductivity
≤0.1μS/cm
≤0.1μS/cm
≤0.1μS/cm
≤0.1μS/cm
≤0.1μS/cm
Resistivity
≥18MΩ
≥18MΩ
≥18MΩ
≥18MΩ
≥18MΩ
Enclosure
Cabinet
Cabinet
Cabinet
Open frame
Open frame
Dimensions
55×35×95 cm
55×35×95 cm
55×35×95 cm
64×64×155 cm
64×64×155 cm
Product Overview
The KY-JY clinical laboratory biochemical water system serves biochemical analyzers and routine diagnostic workflows in 60 / 120 / 180 / 250 and 500 L/h capacities. All five share the same purification process; they differ in output, membrane and polishing-column configuration, enclosure type and the number of analyzers supported. Select the model by sample volume, analyzer count and peak demand.
Key Features
ModelKY-JY-60
Rated output60 L/h
Purification columns2 columns
Warranty1 year system
ModelKY-JY-120
Rated output120 L/h
Purification columns2 columns
Warranty1 year system
ModelKY-JY-180
Rated output180 L/h
Purification columns4 columns
Warranty1 year system
ModelKY-JY-250
Rated output250 L/h
Purification columns6 columns
Warranty1 year system
ModelKY-JY-500
Rated output500 L/h
Purification columns8 columns; expandable to 10
Warranty1 year system
System Design & Functions
Automatic production, full-tank stop, low-pressure protection and membrane flushing.
Purification stages remove ions, residual chlorine, microorganisms and suspended contaminants that can interfere with laboratory analyzers.
KY-JY-60 / 120 / 180 use a closed cabinet with casters; KY-JY-250 / 500 use an open stainless frame — both are easy to install, relocate and service.
Modular components and standardized consumables reduce service complexity.
Designed to support biochemical analyzers, electrolyte analyzers and immunoassay equipment.
Applications
Community hospitals, small outpatient clinics and primary-care laboratories with a single compact biochemical analyzer.
County hospitals and general outpatient laboratories operating up to two biochemical analyzers.
Secondary hospitals and specialist clinics operating three to four biochemical analyzers.
Medium and large hospital laboratories with multiple biochemical and immunoassay systems.
Tertiary hospitals and independent medical laboratories requiring continuous high-flow production.
Installation & Integration
Municipal tap water; stable feed pressure and a nearby drain are required.
Install indoors at 5-38°C away from heat, humidity, sunlight and freezing conditions.
Provide a grounded 220V supply and sufficient front and side service clearance.
Standards & Service References
System water quality and configuration should be confirmed against the connected analyzer requirements.
Each unit is factory checked for water quality, pump pressure and electrical operation.
One-year system warranty; installation, commissioning, operator training and long-term consumable supply are available.